EasyManua.ls Logo

TELEDYNE API T703U - Turning O Generator Calibration Points on; off

Default Icon
273 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Teledyne API T703/T703U Calibrator Operation Manual Calibration and Verification
131
TURNING O
3
GENERATOR CALIBRATION POINTS ON / OFF 6.2.5.
To enable or disable an existing calibration point, press:
STANDBY ACT =STANDBY
<TST TST> GEN STBY SEQ SETUP
Make sure that the
calibrator is in standby
mode.
SETUP X.X PRIMARY SETUP MENU
O3 SEQ CFG CLK PASS MORE EXIT
SETUP X.X SECONDARY SETUP MENU
COMM VARS DIAG EXIT
SETUP X.X ENTER PASSWORD
8 1 8 ENTR EXIT
Toggle to enter the correct
PASSWORD
DIAG SIGNAL I/O
PREV NEXT ENTR EXIT
DIAG O3 GEN CALIBRATION
PREV NEXT ENTR EXIT
Continue pressing NEXT until ...
DIAG O3 GEN CALIBRATION
CAL PNTS EXIT
DIAG O3GEN 1) 500 MV, 5.0 MIN, ON
PREV NEXT INS DEL EDIT PRNT EXIT
Continue pressing PREV & NEXT until your
reach the point to be turned ON/OFF
DIAG O3GEN 8) 1500 MV, 5.0 MIN, ON
PREV NEXT INS DEL EDIT PRNT EXIT
DIAG O3GEN CAL. POINT DRIVE:0 MV
<SET SET> EDIT EXIT
DIAG O3GEN CAL. POINT ENABLELD:ON
<SET SET> EDIT EXIT
Continue pressing SET> until ...
DIAG O3GEN CAL. POINT ENABLELD:ON
ON ENTR EXIT
Turns the cal point ON / OFF
EXIT discards
the new setting
ENTR accepts
the new setting
07223D DCN7107

Table of Contents