Teledyne API T703/T703U Calibrator Operation Manual Calibration and Verification
127
O
3
GENERATOR CALIBRATION TABLE 6.2.1.
When the T703/T703U is operated in BENCH mode, this table is used for the
initial setting only. After a short delay time, the bench feedback control will take
over and control the O
3
generator drive voltage to match the concentration
requested.
The instrument software will interpolate between two values in the table when an
intermediate concentration is requested.
For each point included in the table used by the T703/T703U to calibrate the O
3
generator the user can set a drive voltage and a dwell time for that point. Each
point can also be individually turned off or on.
When the T703/T703U is operated in CONST mode, the initial O
3
generator
drive setting will be set by the calibration table and does not change.
07223D DCN7107