EasyManua.ls Logo

TELEDYNE API T703U - Calibrating Gas Pressure Sensors; Gas Pressure Sensor Calibration Set up; Figure 6-1: Set up for Verifying Optional O Figure 6-2: Set up for Verifying Optional O Figure 6-3: Pressure Calibration Monitor Point - T703; Table 6-1: T703;T703 U Pressure Sensors

Default Icon
273 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Teledyne API T703/T703U Calibrator Operation Manual Calibration and Verification
133
6.3. CALIBRATING GAS PRESSURE SENSORS
The T703 Calibrator has two sensors that monitor the pressure of the gases
flowing through the instrument: a Regulator pressure sensor and a Photometer
Sample Gas pressure sensor. The T703U also has those two sensors plus one
more: an O
3
Generator Regulator pressure sensor. The data collected by these
sensors are used to compensate the final concentration calculations for changes in
atmospheric pressure and are stored in the CPU‟s memory as test functions:
Table 6-1: T703/T703U Pressure Sensors
SENSOR
ASSOCIATED
TEST FUNCTION
UNITS
PRESSURE MONITOR
MEASUREMENT POINT
Regulator Pressure Sensor
REG PRESSURE
PSIG
Capped fitting on backside of regulator
assembly. See Figure 6-5.
Photometer Sample Gas
Pressure Sensor
PHOTO SPRESS
IN-HG-A
Use monitor to measure ambient
atmospheric pressure at the calibrator‟s
location.
(T703U only) O
3
Generator
Regulator Pressure Sensor
O3 PRES
PSIG
Capped fitting on O
3
Generator Regulator
assembly. See Figure 6-6.
GAS PRESSURE SENSOR CALIBRATION SET UP 6.3.1.
The procedures described in this section require an independent, calibrated
pressure meter/monitor to be attached at the location shown in Figure 6-3 for
T703, or at the locations shown in Figure 6-4 for T703U.
Figure 6-3: Pressure Calibration Monitor Point – T703
07223D DCN7107

Table of Contents