Getting Started Teledyne API T703/T703U Calibrator Operation Manual
38
FUNCTIONAL CHECK 3.4.4.
1. After the calibrator‟s components have warmed up for at least 30 minutes,
verify that the software properly supports any hardware options that are
installed.
2. Check to make sure that the calibrator is functioning within allowable
operating parameters. Appendix C includes a list of test functions viewable
from the calibrator‟s front panel as well as their expected values. These
functions are also useful tools for diagnosing problems with your calibrator
(see Section 8.1.2). The enclosed Final Test and Validation Data sheet (part
number 05760) lists these values before the instrument left the factory.
To view the current values of these parameters press the following touch screen
button sequence on the calibrator‟s front panel. Remember until the unit has
completed its warm up these parameters may not have stabilized.
STANDBY ACT =STANDBY
<TST TST> GEN STBY SEQ
2
MSG CLR
1
SETUP
Toggle <TST TST> to
scroll through list of
functions
O3GEN FRAC=[Value] (T703U only)
O3 PRES
3
=[Value] PSIG
A-O3=[Value]
T-O3=[Value]
OUT FLW=[Value] LPM
REG PRES=[Value] PSIG
BOX TMP=[Value]ºC
O3GENDRV=[Value] MV
O3LAMPTMP=[Value]ºC
PH MEAS[Value] MV
PH REF=[Value] MV
PH FLW=[Value] LPM
PH LTEMP=[Value]ºC
PH SPRES=[Value] IN-HG-A
PH STEMP=[Value]ºC
PH SLOPE=[Value]
PH OFFST=[Value] PPB
PH STAB=[Value] PPB
TIME=[HH:MM:SS]
1
Only appears when warning
messages are active.
2
Only appears when one or more
calibration sequences are
programmed into the calibrator‟s
memory.
3
O
3
Generator Regulator Pressure
If your calibrator is operating via Ethernet and your network is running a dynamic
host configuration protocol (DHCP) software package, the Ethernet option will
automatically configure its interface with your LAN. However, it is a good idea
to check these settings to make sure that the DHCP has successfully downloaded
the appropriate network settings from your network server (See Section 5.6.1).
If your network is not running DHCP, you will have to configure the calibrator‟s
interface manually (See Section 5.6.1.1).
07223D DCN7107

Table of Contents