Specifications and Approvals Teledyne API T703/T703U Calibrator Operation Manual
18
Table 2-3: T703/T703U Specifications for O
3
Photometer
Ranges
0-100 ppb to 0-10 ppm, user selectable
Zero Noise
0.3ppb (RMS)
Span Noise
<0.5%
Lower Detectable Limits
0.6 ppb (RMS)
Precision
1.0 ppb
Linearity
1.0% of Full Scale
Lag Time
<10 seconds
Rise/Fall Time
<20 sec (photometer response)
Zero Drift
<1.0 ppb / 7 days
Span Drift
<1% / 24 hours; <2% / 7 days
Flow Rate
800 cc
3
/min +/- 10%
2.2. APPROVALS AND CERTIFICATIONS
The Teledyne API Models T703 and T703U Photometric O
3
Calibrators were
tested and certified for Safety and Electromagnetic Compatibility (EMC). This
section presents the compliance statements for those requirements and directives.
SAFETY 2.2.1.
IEC 61010-1:2001, Safety requirements for electrical equipment for
measurement, control, and laboratory use.
CE: 2006/95/EC, Low-Voltage Directive
EMC 2.2.2.
EN 61326-1 (IEC 61326-1), Class A Emissions/Industrial Immunity
EN 55011 (CISPR 11), Group 1, Class A Emissions
FCC 47 CFR Part 15B, Class A Emissions
CE: 2004/108/EC, Electromagnetic Compatibility Directive
OTHER TYPE CERTIFICATIONS 2.2.3.
For additional certifications, please contact Technical Support: