Teledyne API T703/T703U Calibrator Operation Manual Operating the Calibrator
65
ADJUSTING THE INTERNAL CLOCK’S SPEED 4.7.2.
In order to compensate for CPU clocks which run faster or slower, you can adjust
a variable called CLOCK_ADJ to speed up or slow down the clock by a fixed
amount every day. To change this variable, press:
SETUP X.X 0) O3_PHOTO_LAMP=58.0 DegC
PREV NEXT JUMP EDIT PRNT EXIT
SETUP X.X ENTER SETUP PASS:0
8 1 8 ENTR EXIT
Continue pressing NEXT until ...
SETUP X.X 7) CLOCK_ADJUST=0 Sec/Day
+ 0 0 EDIT ENTR EXIT
SETUP X.X 7) CLOCK_ADJUST=0 Sec/Day
PREV NEXT EDIT ENTR EXIT
Enter sign and number of
seconds per day the clock
gains (-) or loses(+)
SETUP X.X 7) CLOCK_ADJUST=0 Sec/Day
PREV NEXT JUMP EDIT ENTR EXIT
EXIT discards the new
setting
ENTR accepts the
new setting
STANDBY ACT =STANDBY
<TST TST> GEN STBY SEQ SETUP
SETUP X.X PRIMARY SETUP MENU
O3 SEQ CFG CLK PASS MORE EXIT
SETUP X.X SECONDARY SETUP MENU
COMM VARS DIAG EXIT