Teledyne API T703/T703U Calibrator Operation Manual Calibration and Verification
135
Figure 6-6: O
3
Regulator Pressure Monitor Point (T703U only)
CALIBRATING THE PRESSURE SENSORS
6.3.2.
The following procedure requires a pump. If your unit does not have the internal
pump option, temporarily connect an external pump.
1. Turn off the calibrator and open the top cover.
2. Connect a pressure meter to the Regulator Pressure measurement fitting
(see Figure 6-5). This fitting is located on the backside of the regulator
assembly.
3. If the unit is a T703U, also connect a pressure meter to the O
3
Regulator (Figure 6-6).
4. Turn on the calibrator and perform the following steps: