Teledyne API - Models T703, T703U, 703E MENU TREES (05745F DCN7107) APPENDIX A-2: Serial I/O Setup Variables
A-11
APPENDIX A-2: Serial I/O Setup Variables
Table A-1: Setup Variables
Setup Variable
Numeric
Units
Default
Value
Value
Range
Description
PHOTO_LAMP
ºC
58
0100
Photometer lamp temperature
set point and warning limits.
Warnings:
5661
O3_GEN_LAMP
ºC
48
0100
O
3
generator lamp temperature
set point and warning limits.
Warnings:
4353
O3_CONC_RANGE
PPB
500
0.120000
O
3
concentration range for test
channel analog output.
O3_PHOTO_BENCH_ONLY
OFF
OFF, ON
O
3
bench control flag. ON turns
on pump and switches
measure/reference valve only in
bench generation mode.
ZA_PUMP_ENAB
ON
OFF, ON
Zero air pump control. ON turns
on zero air pump when
generating ozone.
STD_TEMP
ºC
25
0100
Standard temperature for unit
conversions.
STD_PRESS
"Hg
29.92
1550
Standard pressure for unit
conversions.
CLOCK_ADJ
Sec./Day
0
-6060
Time-of-day clock speed
adjustment.
SERVICE_CLEAR
OFF
OFF
ON
ON resets the service interval
timer.
TIME_SINCE_SVC
Hours
0
0500000
Time since last service.
SVC_INTERVAL
Hours
0
0100000
Sets the interval between service
reminders.
07223D DCN7107