EasyManua.ls Logo

TELEDYNE API T703U - Status Outputs; Table 8-9: Analog Output Test Function - Nominal Values Voltage Outputs; Table 8-10: Status Outputs Check

Default Icon
273 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
General Troubleshooting & Service Teledyne API T703/T703U Calibrator Operation Manual
170
Table 8-9: Analog Output Test Function - Nominal Values Voltage Outputs
FULL SCALE OUTPUT OF VOLTAGE RANGE
(see Section 4.9.1.3)
100mV
1V
5V
10V
STEP
%
NOMINAL OUTPUT VOLTAGE
1
0
0
0
0
0
2
20
20 mV
0.2
1
2
3
40
40 mV
0.4
2
4
4
60
60 mV
0.6
3
6
5
80
80 mV
0.8
4
8
6
100
100 mV
1.0
5
10
If one or more of the steps fails to be within these ranges, it is likely that there has
been a failure of the either or both of the DACs and their associated circuitry on
the motherboard.
Status Outputs 8.4.9.3.
To test the status output electronics:
1. Connect a jumper between the “D“ pin and the “” pin on the status output
connector.
2. Connect a 1000 ohm resistor between the “+” pin and the pin for the status
output that is being tested.
3. Connect a voltmeter between the “” pin and the pin of the output being
tested (see table below).
4. Under the DIAG> SIGNAL I/O menu (See Section8.1.3), scroll through the
inputs and outputs until you get to the output in question.
5. Alternately, turn on and off the output noting the voltage on the voltmeter.
It should vary between 0 volts for ON and 5 volts for OFF.
Table 8-10: Status Outputs Check
PIN
(LEFT TO RIGHT)
STATUS
1
ST_SYSTEM_OK
2
SPARE
3
ST_CAL_ACTIVE
4
ST_DIAG_MODE
5
ST_TEMP_ALARM
6
ST_PRESS_ALARM
7 and 8
SPARE
07223D DCN7107

Table of Contents