Operating the Calibrator Teledyne API T703/T703U Calibrator Operation Manual
44
4.3. STANDBY MODE
When the T703 Photometric Calibrator is in standby mode, it is at rest.
Some functions under the SETUP > MORE > DIAG submenu, those which
conflict with accurate creation of calibration gas mixtures (e.g. ANALOG
OUTPUT STEP TEST) automatically place the calibrator into STANDBY
mode when activated
NOTE
The T703/T703U calibrator should always be placed in STANDBY mode when not needed to
produce calibration gas.
This can be done manually by pressing the STBY button that appears when the calibrator’s
display is showing the top-level menu (see Figure 4-1).
When programming a calibration sequences the STANDBY step should always be inserted at the
end of the sequence.