Calibration and Verification Teledyne API T703/T703U Calibrator Operation Manual
130
DELETING O
3
GENERATOR CALIBRATION POINTS 6.2.4.
To delete an existing calibration point, press:
STANDBY ACT =STANDBY
<TST TST> GEN STBY SEQ SETUP
Make sure that the
calibrator is in standby
mode.
SETUP X.X PRIMARY SETUP MENU
O3 SEQ CFG CLK PASS MORE EXIT
SETUP X.X SECONDARY SETUP MENU
COMM VARS DIAG EXIT
SETUP X.X ENTER PASSWORD
8 1 8 ENTR
EXIT
Toggle to enter the correct
PASSWORD
DIAG SIGNAL I/O
PREV NEXT ENTR EXIT
DIAG O3 GEN CALIBRATION
PREV NEXT ENTR EXIT
Continue pressing NEXT until ...
DIAG O3 GEN CALIBRATION
CAL PNTS EXIT
DIAG O3GEN 1) 500 MV, 5.0 MIN, ON
PREV NEXT
INS DEL EDIT PRNT EXIT
DIAG O3GEN 8) 1500 MV, 5.0 MIN, ON
PREV NEXT INS DEL EDIT PRNT EXIT
Continue pressing PREV & NEXT until your
reach the point to be deleted
DIAG O3GEN DELETE?
YES
NO
DIAG O3GEN 8) 2000 MV, 5.0 MIN, ON
PREV NEXT INS DEL EDIT PRNT EXIT
DIAG O3GEN DELETED
07223D DCN7107