Operating the Calibrator Teledyne API T703/T703U Calibrator Operation Manual
42
4.1. TEST FUNCTIONS
A variety of TEST FUNCTIONS are available for viewing at the front panel
whenever the calibrator is at the MAIN MENU. These measurements provide
information about the present operating status of the calibrator and are useful
during troubleshooting (see Section 8). Table 6-1 lists the available TEST
functions.
To view these TEST functions, press <TST or TST> from the main menu and
the Test Functions will scroll at the top center of the display.
Table 4-1: Test Functions Defined
TEST MEASUREMENT
DESCRIPTION
O3GEN FRAC=X.XX (T703U only)
For low range mode to produce low levels of ozone during
a GPT calibration where O
3
output is <500 PPB LPM
(supported only in GEN-AUTO mode)
O3 PRES=XX.X PSIG
O
3
generator regulator pressure
ACT=GENERATE XX PPB O3
Actual concentration being generated, computed from
real-time inputs
TARG=GENERATE XX PPB O3
Target concentration to generate
OUT FLW=X.XXX LPM
Output flow rate (computed from regulator pressure)
REG PRESS=XX.X PSIG
Regulator pressure
BOX TMP=XX.X ºC
Internal chassis temperature
O3GENDRV=XXXX.X MV
O
3
generator lamp drive output
O3LAMPTMP=XX.X ºC
O
3
generator lamp temperature
PH MEAS=XXXX.X MV
Photometer detector measure reading
PH REF=XXXX.X MV
Photometer detector reference reading
PH FLW=X.XXX LPM
Photometer sample flow rate
PH LTEMP=XX.X ºC
Photometer lamp temperature
PH SPRES=XX.X IN-HG-A
Photometer sample pressure
PH STEMP=XX.X ºC
Photometer sample temperature
PH SLOPE=X.XXX
Photometer slope computed during zero/span bench
calibration
PH OFFST=X.X PPB
Photometer offset computed during zero/span bench
calibration
PHOTO STABIL=X.X PPB
Photometer concentration stability (standard deviation of
25 bench concentration samples taken 10 seconds apart)
TEST=XXXX.X MV
Value output to TEST_OUTPUT analog output, selected
with TEST_CHAN_ID variable
TIME=HH:MM:SS
Current instrument time of day clock. (24 hour format)