EasyManua.ls Logo

TELEDYNE API T703U - Photometer UV Source Lamp Adjustment

Default Icon
273 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Maintenance Schedule & Procedures Teledyne API T703/T703U Calibrator Operation Manual
148
7.5. PHOTOMETER UV SOURCE LAMP ADJUSTMENT
This procedure details the steps for adjustment of the UV source lamp in the
optical bench assembly. This procedure should be done whenever the PHOTO
REFERENCE test function value drops below 3000 mV.
1. Make sure the instrument is warmed-up and has been running for at least 15
minutes before proceeding.
2. Remove the cover from the instrument.
3. Locate the Photometer (see Figure 3-5)
4. Locate the UV DETECTOR GAIN ADJUST POT on the photometer
assembly (see Figure 7-3).
5. Perform the following procedure:
07223D DCN7107

Table of Contents