1
Only occurs if one of the voltage ranges is selected.
2
Manual adjustment menu only appears if either the AUTO CAL feature is OFF
3
DIAG Menu is inactive while instrument is in GENERATE mode.
MORE DIAG
3
ENTER PASSWORD: 818
SIGNAL I/O
ANALOG
OUTPUT
ANALOG I/O
CONFIGURATION
PHOTO FLOW
SENSOR
CALIBRATION
O3 GEN
CALIBRATION
PRESSURE
CALIBRATION
TEST CHAN
OUTPUT
Press ENTR
to start test
NEXTPREV
0)- 11) CONTROL_IN_1
THRU 12
12)-23) CONTROL_OUT_1
THRU 12
42 INTERNAL ANALOG
to VOLTAGE SIGNALS
ON
OFF
24) ST_SYSTEM_OK
25) ST_CAL_ACTIVE
26) ST_DIAG_MODE
27) ST_TEMP_ALARM
28) ST_PRESS_ALARM
29) RELAY_WATCHDOG
30) ZERO_AIR_PUMP
31) PHOTO_REF_VALVE
32) ZA_SHUTOFF_VALVE
33) O3_PUMP_ON
34) PHOTO_LAMP_HEATER
35) O3_GEN_HEATER
36) MAINT_MODE
37) LANG2_SELECT
38) SEQUENCE_LED
39) AUTO_TIMER_LED
40) FAULT_LED
41) AUDIBLE_BEEPER
AOUTS CALIBRATED AIN CALIBRATED
SET><SET
CAL
Initiates auto-calibration of all
analog outputs
Initiates auto-calibration of the analog inputs’
zero and span points
CAL
EDIT
RANGE
10V
5V1V0.1V
ON
OFF
OVER
RANGE
REC
OFFSET
1
AUTO
1
CAL
ON
OFF
CAL
CALIBRATED
Sets a voltage
offset for the
output
Auto Cal
1
U100 UP10 UP DOWN DN10 D100
Manual Cal
2
Initiates auto-calibration
of the selected analog input
See SECONDARY SETUP Menu
DIAG - CALIBRATION Submenu
PREV NEXT
PREV NEXT
MAIN MENU
SETUP
CONC_OUT_1
CONC_OUT_2
TEST_OUTPUT
SET><SET
O3GEN FRAC=[Value] (T703U only)
O3 PRES=[Value] PSIG
A-O3=[Value]
T-O3=[Value]
OUT FLW=[Value] LPM
REG PRES=[Value] PSIG
BOX TMP=[Value]ºC
O3GENDRV=[Value] MV
O3LAMPTMP=[Value]ºC
PH MEAS[Value] MV
PH REF=[Value] MV
PH FLW=[Value] LPM
PH LTEMP=[Value]ºC
PH SPRES=[Value] IN-HG-A
PH STEMP=[Value]ºC
PH SLOPE=[Value]
PH OFFST=[Value] PPB
PH STAB=[Value] PPB
TIME=[HH:MM:SS]