Operating the Calibrator Teledyne API T703/T703U Calibrator Operation Manual
52
To set the delta timer, press:
STANDBY ACT =STANDBY
<TST TST> GEN STBY SEQ SETUP
Make sure that the
calibrator is in standby
mode.
SETUP X.X PRIMARY SETUP MENU
O3 SEQ CFG CLK PASS MORE EXIT
SETUP X.X SEQUENCE CONFIGURATION
EDIT PRINT EXIT
SETUP X.X END OF SEQUENCES
INS PRNT EXIT
SETUP X.X 1) SEQ [NAME], [X] STEPS
PREV NEXT INS DEL EDIT PRNT EXIT
SETUP X.X NAME:0
SET> EDIT EXIT
Deletes the sequence shown
in the message field
Edits the sequence shown
in the message field
Continue pressing SET> until ...
This display only appears if there are no sequences currently
programmed into the calibrator.
OTHERWISE ...
SETUP X.X TIMER DELTA: 001:00:00
<SET SET> EDIT EXIT
SETUP X.X TIMER DELTA: 0 Days
0 0 0 ENTR EXIT
Toggle these buttons to
enter number of days to
wait between before
running sequence again.
EXIT discards the
new setting
ENTR accepts the
new setting
SYSTEM TIMER DELTA 00:00
1 2 :0 0 ENTR EXIT
HOUR MINUTE
Toggle these
buttons to enter the
starting time
EXIT discards the
new setting
ENTR accepts the
new setting
Scrolls back and forth between
existing sequences
07223D DCN7107