EasyManua.ls Logo

TELEDYNE API T703U - Page 87

Default Icon
273 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Teledyne API T703/T703U Calibrator Operation Manual Operating the Calibrator
85
To use the T703/T703U as a photometer, press:
STANDBY ACT =STANDBY
<TST TST> GEN STBY SEQ SETUP
BCAL XZRO ACT=403.4 PPB O3
<TST TST> ZERO SPAN EXIT
STANDBY ACT =STANDBY
<TST TST> GEN STBY SEQ SETUP
SETUP X.X PRIMARY SETUP MENU
O3 SEQ CFG CLK PASS MORE EXIT
SETUP X.X O3 GAS CONFIG
MODE ADJ PHOT EXIT
SETUP X.X O3 GAS CONFIG
BCAL DARK EXIT
SYSTEM ENTER PASSWORD:0
0 0 0 ENTR EXIT
Toggle these buttons until
PASSWORD = 717
SETUP X.X O3 PHOTOMETER BENCH CAL
EDIT CAL EXIT
SETUP X.X BENCH CAL:XZRO
ZERO SPAN XZRO XSPN ENTR EXIT
Press XZRO if using an
External O
3
Generator
The ACT function in the
message field displays the
O
3
concentration of the
source gas
SETUP X.X O3 PHOTOMETER BENCH CAL
EDIT CAL EXIT
Make sure that the
calibrator is in STANDBY
mode.
DO NOT!
Press the GEN button
Make sure that the
ACT test function is
displayed in the
message field
DO NOT!
Press the ZERO or
SPAN buttons. Doing so
will cause the instrument
to recalculate it internal
SLOPE and OFFSET
values.
When finished measuring the external O
3
source, press EXIT.
Continue pressing EXIT until ...
07223D DCN7107

Table of Contents