Sequencing settings (base calling)
The default seings are optimized for sequencing of PCR amplicons from diploid
genes, which are expected to have Mixed bases and which should end At PCR Stop.
For sequencing from plasmid templates, which are of pure sequence (no Mixed bases)
and have longer sequence read times than PCR products, create new analysis seings
and disable (uncheck) At PCR Stop and Mixed base threshold checkboxes.
Setting Description
Quality Threshold
Basecall assignment (ambiguous bases):
Do not assign Ns to basecalls
Assign Ns to basecalls with QV<5— Bases with a QV less than the threshold
display N instead of the base letter
End base—Last base on which to perform basecalling:
At PCR Stop
After X number of bases
After X number of Ns in X number of bases
After X number of Ns
Note: If you have PCR products with sequences that end while data is still being
collected, select the At PCR Stop checkbox.
Mixed bases threshold
When enabled, determines the secondary peak height ratio where the secondary peak
is considered a potential mixed base. Reaching the threshold is a necessary but not
sufficient condition for the basecalling algorithm to call a mixed base.
Clear range methods
Use quality values—Sets a window with a specified number of allowed low-quality
bases by removing bases until there are <X number of bases per Z number of
bases with QV <Y.
Use base positions—Specifies the first and last base in the range to consider, or
trims the specified number of bases from the 3¢ end.
Mask base positions before—Specifies the base position before which to disregard
bases.
The default le name convention determines how the data les (AB1 or FSA)
associated with a plate are named.
The default le name convention is:
Application
Default settings
Fragment/HID analysis <well>_<sample name>_<sample type>_<date and timestamp>.fsa
Sequence analysis <well>_<sample name>_<date and timestamp>.ab1
Sequence analysis with the
Sanger variant analysis
option selected
<well>_<sample name>_<amplicon>_<specimen>_<date and
timestamp>.ab1
Modify the default
file name
convention
Chapter 10 Manage the software (instrument)
Manage run settings (instrument)
10
146
SeqStudio
Genetic Analyzer Instrument and Software User Guide

Table of Contents

Other manuals for Thermo Scientific Applied Biosystems SeqStudio Genetic Analyzer

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Thermo Scientific Applied Biosystems SeqStudio Genetic Analyzer and is the answer not in the manual?

Thermo Scientific Applied Biosystems SeqStudio Genetic Analyzer Specifications

General IconGeneral
BrandThermo Scientific
ModelApplied Biosystems SeqStudio Genetic Analyzer
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals