The SeqStudio
™
Genetic Analyzer Cathode Buer Container (CBC) contains running
buer for capillary electrophoresis. The container has two compartments. The rear
compartment provides the cathode buer for electrophoresis. The front compartment
is for capillary wash and waste.
The CBC requires a Reservoir Septa.
See “Assemble the SeqStudio
™
Genetic Analyzer Cathode Buer Container (CBC)“ on
page 166.
Figure 7 SeqStudio
™
Genetic Analyzer Cathode Buffer Container with Reservoir Septa
1
Reservoir Septa
2
Notch (inserted towards the rear right in the autosampler)
3
Fill line (replace the CBC when the buffer is at the fill line)
4
Cathode buffer compartment
5
Waste and wash compartment
Figure 8 Autosampler
1
Location of plate or tubes
2
Release button
3
Location of Cathode Buffer Container
4
Location to position the Cathode Buffer Container notch
The cartridge and cathode buer have radio–frequency identication (RFID) tags. The
instrument reads and tracks:
• Expiry date (shelf life)
• Remaining injections (usage)
• Serial number (cartridge only)
• Lot number (cathode buer only)
SeqStudio
™
Genetic Analyzer
Cathode Buffer
Container
Radio frequency
identification
Chapter 1 Product information
SeqStudio
™
Genetic Analyzer consumables
1
20
SeqStudio
™
Genetic Analyzer Instrument and Software User Guide