This function manages the list of size standard denitions that you can select from
during plate set up.
To assign a size standard denition to a plate, see “Assign wells: run module, size
standard, dye set, and kit“ on page 84.
A size standard denes the sizes in basepairs of known fragments. It is used to
generate a standard curve. The standard curve is used to determine the sizing of
fragments in unknown samples.
Factory-installed items cannot be edited or deleted. To create a new item from a
factory-installed item, copy, edit, then save the new item.
Size standard denitions are accessible to all users.
1.
Access the Manage size standards screen:
From Actions
Plate properties tab Touch More options4Manage size standards.
Home screen Touch Settings4Run settings 4Size standards.
2.
To create a new size standard:
a.
Touch a default size standard or a user-created size standard, then touch
Copy.
b.
Enter a name and select a dye color.
c.
As needed, touch any of the following, then touch Done:
Note: To change an existing value, add a new value, then delete the original
value.
One or more fragment size values, then touch Delete.
Add to add a value.
3.
As needed, touch a size standard of interest, then touch Edit or View.
Edit fragment/HID analysis settings
Factory-installed items cannot be edited or deleted. To create a new item from a
factory-installed item, copy, edit, then save the new item.
Analysis seings are accessible only to the user who creates them.
1.
Access the Analysis seings screen:
From Action
Plate properties tab Touch More options4Analysis settings, then select the
settings to assign to the plate setup.
Home screen Touch Settings4Run settings 4Analysis settings, then
manage the list of settings that you can select from during
plate set up.
2.
To create new analysis seings:
a.
Touch the default analysis seings or user-created analysis seings, then
touch Copy.
Manage size
standard
definitions
Manage analysis
settings
Chapter 10 Manage the software (instrument)
Manage run settings (instrument)
10
142
SeqStudio
Genetic Analyzer Instrument and Software User Guide

Table of Contents

Other manuals for Thermo Scientific Applied Biosystems SeqStudio Genetic Analyzer

Related product manuals