Three types of calibration can occur on the instrument:
Factory calibration—Default calibration provided with the instrument. It is not
optimized for a specic instrument.
Manual calibration—Manual procedure performed by the user that provides a
baseline calibration for the instrument on which it is run. Reduces pull-up (false
secondary peaks under a true peak).
Auto calibration—Automatic adjustment of the baseline calibration to optimally
reduce pull-up (false secondary peaks under a true peak). The instrument
performs an auto calibration for system and custom dyes.
Sequence analysis dye sets—Do not typically require manual calibration.
Fragment analysis dye sets—Manual calibration is recommended one time
before use.
Note: With high-quality sample data, it is possible for auto calibration to pass
using the factory calibration.
HID analysis dye sets—Manual calibration is required for each dye set one time
before use.
To determine if a dye set requires manual calibration, review the calibration history
for the dye set.
In the home screen:
Factory, auto, and
manual
calibration
Determine if
manual
calibration is
required
Appendix C Dye calibration and install standard checks
Calibrate dyes
C
SeqStudio
Genetic Analyzer Instrument and Software User Guide
193