SeqStudio
Integrated Capillary Protector storage
Condition
Description
Shipping Is shipped at 2–8°C.
Before opening, can be stored until expiry date on label at 2–8°C.
Discard the shipping ICP when you insert a new cartridge in the instrument.
Reuse Do not reuse.
Use a new ICP for off-instrument cartridge storage.
IMPORTANT! If an ICP is used for off-instrument cartridge storage more than
one time, the capillary array will not be kept fully hydrated.
You can leave the cartridge installed in the instrument if the instrument is powered on
and a CBC is installed. Replace the CBC every 2 weeks.
IMPORTANT! If the instrument will be idle for longer than the shelf life of the
cartridge components, remove the cartridge and cathode buer, then power o the
instrument.
Table 12 Cartridge storage
Cartridge
Storage
SeqStudio
Genetic Analyzer Cartridge
(Cat. No. A33671)
See Table 1 on page 18
SeqStudio
Genetic Analyzer Cartridge v2
(Cat. No. A41331)
See Table 2 on page 19
IMPORTANT! The instrument does not maintain the correct temperature conditions
for the cartridge when it is powered o. Avoid cartridge exposure to ambient
temperature.
Long-term on-
instrument
cartridge storage
Chapter 11 Maintain the instrument
Manage the cartridge
11
SeqStudio
Genetic Analyzer Instrument and Software User Guide
165

Table of Contents