Create and run an HID plate
HID analysis requirements ........................................... 212
Pull-up reduction feature ............................................. 213
Resolution algorithm ................................................ 214
Workow: HID analysis .............................................. 215
HID install run ...................................................... 215
HID run module, dye sets, size standard denitions, and kits ............. 216
Manage kits (HID applications only) ................................... 217
HID analysis requirements
Cartridge—Use SeqStudio
Genetic Analyzer Cartridge v2 (Cat. No. A41331),
which has been validated for HID analysis. The v1 cartridge has not been
validated.
To check the Cat. No., you must remove the cartridge from the instrument. See
“Store the cartridge“ on page 163 for instructions.
Manual dye calibration and HID install check—When the instrument is
installed, the eld service representative will perform a manual J6 dye calibration
and an HID install check.
Before you use any other dye sets, perform a manual dye calibration to optimize
the pull-up reduction function. For more information, see “Calibrate dyes“ on
page 192.
Allelic ladder—We recommend that you inject one allelic ladder for each set of
24 samples in HID runs (1 allelic ladder per 6 injections).
Allelic ladders that are injected under the same conditions are recommended to
accurately genotype samples in the secondary analysis software.
IMPORTANT! Variation in laboratory temperature can cause changes in fragment
migration speed that can, in turn, cause sizing variation. We recommend the
frequency of allelic ladder injections described above to account for normal
variation in fragment migration speed. However, during internal HID validation
studies, verify the required allelic ladder injection frequency to ensure accurate
genotyping of all samples in your laboratory environment.
GeneMapper
ID-X Software compatibility—The data les that are generated
by the instrument are compatible with GeneMapper
ID-X Software v1.6 or later.