Monitor a run from the instrument
In the Plate properties screen:
1.
Touch Injection options.
2.
Touch an injection group, then congure the injection list:
Touch and drag an injection group to a new location in the injection list.
Touch Inject rst—Moves the selected injection group to the top of the
injection list.
Touch Edit and re-inject—Adds replicates or re-injections to the injection list.
You can also modify Run module, Injection time, Injection voltage, Run
time, or Run voltage for these injections.
Note: Replicate indicates a duplicate—the replicate has the same parameters
as the original injection. Re-injection indicates a copy of the injection, but with
user-modied parameters.
Note: The changes to the run conditions apply only to the replicate or re-
injection. The changes are not saved to the run module.
3.
Touch Done.
During a run, the run status screen is displayed.
View the run time information and the status dial for each capillary.
The status dials are color-coded for quality alerts:
—All QC tests passed.
—At least 1 warning quality alert was triggered.
—At least 1 failing quality alert was triggered.
If an injection group is set to re-inject, the number of the current injection is displayed
on the status dials.
For information on quality alerts, see:
“Data quality alerts“ on page 119
“Sizecalling and basecalling quality alerts“ on page 119
(Optional)
View
the injection list,
change injection
settings or order,
or specify
replicates and re-
injections
View the run
status
Chapter 7 Start and monitor a run
Monitor a run from the instrument
7
SeqStudio
Genetic Analyzer Instrument and Software User Guide
105

Table of Contents

Other manuals for Thermo Scientific Applied Biosystems SeqStudio Genetic Analyzer

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Thermo Scientific Applied Biosystems SeqStudio Genetic Analyzer and is the answer not in the manual?

Thermo Scientific Applied Biosystems SeqStudio Genetic Analyzer Specifications

General IconGeneral
BrandThermo Scientific
ModelApplied Biosystems SeqStudio Genetic Analyzer
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals