HID analysis install run pass/fail criteria
Result
Acceptance criteria
# Allele peaks 343 for all four capillaries
# Size standard peaks 26 for all four capillaries
Minimum peak height >400 RFU
Standard deviation of the observed allele fragment sizes
for all 4 capillaries
<0.15 base pairs (bp)
Sequence analysis install run pass/fail criteria
Result
Acceptance criteria
QV30 CRL (contiguous read
length)
Note: The contiguous read
length passing criteria for
install checks is an
uninterrupted segment of
bases with an average Quality
Value (QV) of 30.
BigDye
™
Terminator v3.1
Capillaries with QV30 CRL ≥500 bp pass.
BigDye
™
Terminator v1.1
Capillaries with a QV30 CRL ≥500 bp pass.
The remaining results on the screen are for information only.
1.
Touch Seings4Maintenance and Service4Install run4Install run history.
2.
Touch the install run of interest.
3.
Touch Export report.
4.
Select a destination for the report.
5.
Touch OK.
Export the install
run report