Cartridge storage
Table 1 Storage information for the SeqStudio
Genetic Analyzer Cartridge (Cat. No. A33671)
Condition Description
Shipping Shipped at 2–8°C.
Store upright at 2–8°C upon receipt.
Save the white storage box and optical cover for off-instrument cartridge storage.
On-instrument storage For routine use, can be used and stored on the instrument for up to 4 months. If
you store the cartridge on-instrument:
The instrument must be powered on.
A Cathode Buffer Container must also be installed.
The instrument keeps the components under the following conditions when it is
powered on and in Cartridge storage mode:
Optical detection window—Covered
Capillary array electrodes—Submerged in cathode buffer (buffer must be
above the fill line in the Cathode Buffer Container)
Polymer—Chilled
Anode buffer—Ambient temperature
IMPORTANT! The instrument does not maintain the correct temperature
conditions for the cartridge when it is powered off. Avoid cartridge exposure to
ambient temperature.
Off-instrument storage For intermittent use, can be stored off-instrument until the expiry date on the label
or up to 4 months after first use. Store upright at 2–8°C, with an integrated
capillary protector (ICP) and optical cover installed (see “Store the cartridge“ on
page 163).
Note: After you remove the cartridge from the instrument, install an ICP within a
few minutes. Avoid cartridge exposure to ambient temperature.
Reuse Can be removed from an instrument then inserted again on the same instrument
or a different instrument, if it was stored properly at 2–8°C and has not expired or
exceeded 125 injections.
Information about the cartridge installation and usage is retained in the cartridge
history (Settings4Cartridge4Instrument–cartridge history).
Chapter 1 Product information
SeqStudio
Genetic Analyzer consumables
1
18
SeqStudio
Genetic Analyzer Instrument and Software User Guide

Table of Contents

Other manuals for Thermo Scientific Applied Biosystems SeqStudio Genetic Analyzer

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Thermo Scientific Applied Biosystems SeqStudio Genetic Analyzer and is the answer not in the manual?

Thermo Scientific Applied Biosystems SeqStudio Genetic Analyzer Specifications

General IconGeneral
BrandThermo Scientific
ModelApplied Biosystems SeqStudio Genetic Analyzer
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals