1.
Touch Seings4Maintenance and Service4Calibration4Calibration
history.
2.
Examine the entry for the dye set of interest. If the dye set does not list a date and
a cartridge serial number for a manual calibration or a date for auto calibration,
the dye set requires manual calibration.
2
4
3
1
1
5
1
Figure 21 Fragment analysis dye sets example
1
Manual calibration is recommended—No auto calibration or manual calibration has
been performed.
2
Manual calibration is not required—A manual calibration has been performed.
3
Manual calibration is not required—The factory calibration has been optimized for the
instrument by auto calibration.
4
Manual calibration is recommended—Auto calibration has been performed, but only
2 capillaries passed.
5
Manual calibration is not required—A manual calibration and an auto calibration have
been performed.
Appendix C Dye calibration and install standard checks
Calibrate dyes
C
194
SeqStudio
Genetic Analyzer Instrument and Software User Guide