TH2829X Series Operation Manual Chapter 7 Transformer Auto Scanning Test
94
TH2829x716
As shown above, the SCAN PAGE FROM is 2, SCAN PAGE NUMS is 03. It means the
test pages are 2, 3, 4, totally 3 pages.
NOTE: when 0 is entered in SCAN PAGE FROM, it means to turn off this function
and the SCAN PAGE NUMS does not work (defaults to OFF).
7.7.14 RESCAN IF FAIL
On <TOOLS> page, move the cursor to RESCAN IF FAIL and press the number to
confirm the rescan times if fails. It ranges from 1~9. When 0 is entered, this
function is turned off.
TH2829x717
NOTE: the time interval of RESCAN IF FAIL can be set in RESCAN INTERVAL in
<TRANSFORMER ID> page.