TH2829X Series Operation Manual Chapter 7 Transformer Auto Scanning Test
TH2829x716
As shown above, the SCAN PAGE FROM is 2, SCAN PAGE NUMS is 03. It means the
test pages are 2, 3, 4, totally 3 pages.
NOTE: when “0” is entered in SCAN PAGE FROM, it means to turn off this function
and the SCAN PAGE NUMS does not work (defaults to OFF).
7.7.14 RESCAN IF FAIL
On <TOOLS> page, move the cursor to RESCAN IF FAIL and press the number to
confirm the rescan times if fails. It ranges from 1~9. When “0” is entered, this
function is turned off.
TH2829x717
NOTE: the time interval of RESCAN IF FAIL can be set in RESCAN INTERVAL in
<TRANSFORMER ID> page.