TH2829X Series Operation Manual Chapter 10 HANDLER
0 (decimal 48) is equal to OFF.
For example: WrtCmd (“TRAN:DCR:STAT:ON”)
Query syntax: TRAN:DCR:STAT?
Return format: <NR1><NL^END>
The TRAN:DCR:LIMit command is used to set the DCR nominal value, high and
low limit. The TRAN:DCR:LIMit? query returns the current DCR nominal value,
high and low limits.
Command syntax: TRAN:DCR:LIMit <value>,<low limit>,<high limit>
Where,
<value> is the DCR nominal value in NR1, NR2 or NR3 data format followed by Ω.
<low limit> is the DCR low limit in NR1, NR2 or NR3 data format.
<high limit> is the DCR high limit in NR1, NR2 or NR3 data format
NOTE: The low limit should be smaller than the high limit, or error
information will be reported.
For example: WrtCmd(“TRAN:DCR:LIMIT 50 Ω, -0.01, 0.01”)
Query syntax: TRAN:DCR:LIMI?
Return format: <NR3>,<NR3><NR3><NL^END>
The TRAN:MODE is used to set the transformer working mode. The TRAN:MODE?
query returns the current working mode of the transformer.
Command syntax:
TRAN:MODE SEQuence
STEPped
Where,
SEQuence refers to sequential mode.
STEPped refers to single step mode.
For example: WrtCmd (“TRAN:MODE SEQ”)
Query syntax: TRAN:MODE?
Return format: SEQ
<NL^END>
STEP
9.2 GPIB Common Commands
●*RST ●*TRG ●*IDN ●*TST
●*ESE ●*SRE ●*ESR ●*STB
●*OPC ●*CLS
The *RST command resets the instrument.
For example: WrtCmd (“*RST”)
The *TRG command triggers the measurement and then sends the result to