TH2829X Series Operation Manual Chapter 10 HANDLER
187
Return format: <BIN1 count>, <BIN2 count>, , <BIN9 count>, <OUT OF BIN
count>, <AUX BIN count><NL^END>
Where,
<BIN1-9 count> is the count result of BIN1-9, in NR1 data format.
<OUT OF BIN count> is the count result of the OUT OF BIN, in NR1 data format.
<AUX BIN count> is the count result of the auxiliary bin, in NR1 data format.
The COMParator:BIN:COUNt:CLEar command is used to clear all bin count
results.
Command syntax: COMParator:BIN:COUNt:CLEar
For example: WrtCmd (COMP:BIN:COUN:CLE)
9.1.16 Mass MEMory subsystem commands
The Mass MEMory subsystem commands are used for file storing and load.
Command tree:
Mass MEMory :LOAD :STATe <record number>
:STORe :STATe <record number>,<string>
The MMEMory:LOAD:STATe command is used to load the existed file.
Command syntax: MMEMory:LOAD:STATe<value>
Where,
<value> is the file number ranging from 0 to 39 (NR1).
For example: WrtCmd (MMEM:LOAD:STAT 1)
The MMEMory:STORe:STATe command is used to storing the current setting to a
file.
Command syntax: MMEMory:STOR:STATe<value>, <string>
Where,
<value> is the file number ranging from 0 to 39 (NR1).
<string> can be ASCII character string (maximum length is 16).
For example: WrtCmd (MMEM:STOR:STAT 1, Resistor meas)
or WrtCmd (MMEM:STOR:STAT 1), IF ,<string> has not been
input, the default file name will be stored.
9.1.17 TRAN subsystem commands
The TRAN subsystem commands are mainly used to transformer parameters,
such as turn-ratio, primary inductance, leakage inductance, stray capacitance,
impedance, AC impedance, DC resistance, working mode and limits of each
parameter.