TH2829X Series Operation Manual Chapter 3 Introduction to [MEAS]
especially useful.
Operation steps for setting the compare function ON/OFF
1) Move the cursor to COMP, the following soft keys will be displayed.
ON
OFF
2) Use above soft keys to set the compare function as ON or OFF.
3.8.5 Auxiliary bin ON/OFF
When it is necessary to sort the secondary parameters, the limits of the
secondary parameter can be set in HIGH
and LOW of 2
nd
.
Three cases may occur in the process of secondary parameter sorting:
On <LIMIT TABLE SETUP> page, no low / high limit of the secondary
parameters has been set.
On < LIMIT TABLE SETUP > page, the low/high limit of secondary
parameters has been set but Aux
function is set as OFF.
In this case, only those components whose secondary parameters are qualified
can perform primary parameter sorting according to sorting limits. If the
secondary parameters are unqualified and the corresponding primary parameters
are within limit ranges, those components will be sorted into BIN OUT.
On <Limit list setup> page, the low/high limit of the secondary parameters
has been set and the Aux
function is set as ON.
If the primary parameter is out of the limit range, it is sorted into BIN OUT. If the
primary parameter of DUT is within the limit range but its secondary parameter is
out of the limit range, the DUT will be sorted into the Aux
bin.
———————————————————————————————————————
Note: When the secondary parameter only has low limit and the auxiliary bin is
set as ON, if the primary parameter of DUT is within the limit range and the
secondary parameter is smaller than or equal to its low limit, the DUT will be
sorted into the auxiliary bin. When the secondary parameter only has high limit
and the auxiliary bin is set as ON, if the primary parameter of DUT is within the
limit range and the secondary parameter is larger than or equal to its high limit,
the DUT will be sorted into the auxiliary bin.
———————————————————————————————————————
Operation steps for setting the auxiliary bin function ON/OFF
1) Move the cursor to AUX, the following soft keys will be displayed.
ON
OFF
2) Use above soft keys to set the auxiliary function as ON or OFF.
3.8.6 HIGH/LOW
TH2829X can set bin limits of 9 primary parameters and one secondary parameter.