TH2829X Series Operation Manual Chapter 3 Introduction to [MEAS]
44
The test results can be sorted into 10 bins at most (BIN 1 to BIN 9 and BIN OUT).
The high/low limits of primary parameters can be set in high limit and low limit of
bins from BIN 1 to BIN 9. The limit of the secondary parameter can be set in
HIGH
and LOW of 2
nd
.
Operation steps for setting high/low limit
Execute the following steps to set sorting limits.
1) Set PARAM and NOM in the compare function menu and the limit MODE of
the primary parameter.
2) Move the cursor to Low limit of BIN 1. If you select tolerance mode, the
following operation steps should be from step 3 to step 6; if you select
sequential mode, the following operation steps should be from step 7 to step
11.
3) User numeric keys to input low limit value in Low limit. After inputting the
data, you can use (p, n, µ, m, k, M, *1) to replace [ENTER] to input the limit
value. When [ENTER] is used, the default unit is the same as that being input
last time. When pressing *1, the default unit will be F, H or Ω. After inputting
limit value in LOW
of BIN 1, the low limit of BIN 1 will be automatically set as
(absolute limit) and the high limit will be + (absolute limit).
4) The cursor will automatically move to LOW of BIN 2. Repeat step 4 until the
limits of BIN 9 are input. Then the cursor will automatically move LOW
of 2
nd
.
5) After inputting the low limit of the secondary parameter, the cursor will
automatically move to HIGH
of 2
nd
.
6) Input the high limit of the secondary parameter.
7) In Low limit of BIN 1, use numeric keys to input the low limit. After inputting
the data, you can use (p, n, µ, m, k, M, *1) to replace [ENTER] to input the
limit value. When [ENTER] is used to input the limit value, the default unit is
the same as that being input last time. When pressing *1, the default unit will
be F, H or Ω.
8) After inputting the low limit of BIN 1, the cursor will automatically move to
HIGH
of BIN 1. Input the high limit of BIN 1.
9) The cursor will automatically move HIGH of BIN 2. For the limit mode is
sequential mode, the low limit of BIN 2 will be the high limit of BIN 1. Input
the high limit of BIN 2.
10) Repeat step 9 until the high limit of BIN 9 is input. Then the cursor will
automatically move to LOW
of 2
nd
. Input the low limit of the secondary
parameter.
11) The cursor will automatically move to HIGH of 2
nd
. Input the high limit of the
secondary parameter.
3.9 <LIST SWEEP SETUP>
Press [SETUP] and then LIST SWEEP to enter into the <LIST SWEEP SETUP>
page as shown below.