TH2829X Series Operation Manual Chapter 8 Performance and Test
20Hz~500kHz(TH2829B/BX))
20Hz~1MHz (TH2829C/CX)
Min. resolution:1mHz
8.2.2 Signal mode
Normal: When testing, on measurement display page, voltage across test
terminals may be smaller than preset voltage.
Constant level: The auto adjustment of internal level makes the voltage of DUT
accordant with preset voltage.
8.2.3 Test signal level
5mV
RMS
—10V
RMS
10mV
RMS
—5V
RMS
±(10%×preset value+2mV)
±(6%×preset value+2mV)
5OμA
RMS
—100mA
RMS
100μA
RMS
—50mA
RMS
± ( 10%×preset value+
10μA
RMS
)
± ( 6%×preset value+
10μA
RMS
)
When testing Turns Ratio, Phase, the voltage range is : 5mVRMS—10VRMS, ±
(10%×preset value+2mV), with resolution of 1mV.
8.2.4 Output impedance
30Ω, 50Ω and 100Ω ±2% are selectable.
8.2.5 Monitor for test signal level
5mV
RMS
—2V
RMS
0.01mV
RMS
—5mV
RMS
±(3%×reading+0.5mV)
±(12%×reading +0.1mV)
50μA
RMS
—20mA
RMS
0.001μA
RMS
—50μA
RMS
±(3%×reading +5μA)
±(12%×reading +1μA)
8.2.6 Maximum measurement display range
Measurement display range