TH2829X Series Operation Manual Chapter 10 HANDLER
internal resistor mode. The Output RESistor? query returns the current output
internal resistance status.
Command syntax:
30
ORESistor 50
100
For example: WrtCmd (“ORES 30”) Set the output internal resistance is 30 OHM.
Query syntax: ORESistor?
Return format: <NR1><NL^END>
9.1.7 OUTPut subsystem commands
The OUTPut subsystem commands are mainly used to set the status of the 1A DC
bias source and DC isolation function.
Command tree:
OUTPut :High POWer ON(1) (valid only when TH2901-1A is installed)
OFF(0)
:DC :ISOLation ON (1)
OFF (0)
The OUTPut:HPOWer command is used to set the 1A DC bias source as ON or OFF.
The OUTPut:HPOWer? query returns the current status of the 1A DC bias source.
Before using this command, please ensure the option component 1A DC bias
source has been installed.
Command syntax:
ON
OUTPut:HPOWer OFF
1
0
Where,
Character 1 (49) is equal to ON.
Character 0 (48) is equal to OFF.
Query syntax: OUTPut:HPOWer?
Return format: <INT><NL^END> or <OPT1><NL^END>, OPT means the 1A