TH2829X Series Operation Manual Chapter 10 HANDLER
161
Query syntax: VOLTage?
Return format: <NR3><NL^END>
9.1.4 CURRent subsystem commands
The CURRent subsystem commands are mainly used to set the measurement
current. The CURRent? query returns the current measurement current.
Command syntax:
<value>
CURRent MIN
MAX
Where,
<value> NR1, NR2 or NR3 data format followed by MA.
MIN Set the measurement current as 50μA.
MAX Set the measurement current as 20mA.
For example: WrtCmd (CURR 10MA) Set the measurement current as 10mA.
Query syntax: CURRent?
Return format:<NR3><NL^END>
9.1.5 AMPLitude subsystem commands
The :AMPLitude subsystem commands are mainly used to set the auto level
control (ALC) function as ON or OFF. The :AMPLiture:ALC? query returns the
current status of the ALC function.
Command systax:
ON
AMPLitude:ALC OFF
1
0
Where,
Character 1 (49) is equal to ON.
Character 0 (48) is equal to OFF.
For example: WrtCmd (AMPL:ALC 0) Set the ALC function as OFF.
Query syntax: AMPLitude:ALC?
Return format: <NR1><NL^END>
9.1.6 Output RESistor subsystem commands
The Output RESistor subsystem commands are mainly used to set the output