TH2829X Series Operation Manual Chapter 10 HANDLER
173
9.1.12 TRIGger subsystem commands
The TRIGger subsystem commands are mainly used to set the instrument trigger
source, trigger delay and trigger measurement.
Command tree:
TRIGger [:IMMediate]
:SOURce INTernal
EXTernal
BUS
HOLD
:DELay <value>
MIN
MAX
The TRIGger[:IMMediate] command is used to trigger a test.
Command syntax: TRIGger[:IMMediate]
For example: WrtCmd(TRIG)
The TRIGger:SOURce command is used to set the trigger source mode. The
TRIGger:SOURce? query returns the current trigger source mode.
Command syntax:
INTernal
TRIGger:SOURce EXTternal
BUS
HOLD
Where,
INTernal The default trigger mode.
EXTernal Triggered by HANDLER interface.
BUS Triggered by RS232C interface or GPIB interface
HOLD Triggered by pressing TRIGGER.
For example: WrtCmd (TRIG:SOUR BUS)
Query syntax: TRIGger:SOURce?
Return format:
INT
EXT <NL^END>
BUS
HOLD
The TRIGger:DELay command is used to set the delay time after triggering. The
TRIGger:DELay? query returns the current delay time.