TH2829X Series Operation Manual Chapter 10 HANDLER
B Compare the secondary parameter of the test results with
the high and the low limits.
OFF No comparison
<low limit n> NR1, NR2 or NR3 data format, low limit of the sweep point
on the n
th
line.
<high limit n> NR1, NR2 or NR3 data format, high limit of the sweep
point on the n
th
line.
For example: WrtCmd (“LIST:BAND1 A, 10, 20”)
WrtCmd (“LIST:BAND3 OFF”)
Query syntax: LIST:BAND<n>?
Return Format: <parameter>,<low limit n>,<high limit n>
The LIST:CLEar:ALL command is used to clear the setup data of all the sweeping
points.
Query syntax: LIST:CLEar:ALL
9.1.11 APERture subsystem commands
The APERture subsystem commands are mainly used to set the measurement
speed, average times used in measurement. The APERture? query returns the
current measurement speed, average times.
Command syntax:
FAST
APERture MEDium [,<value>]
SLOW
Where,
FAST: 30 times/sec
MEDium: 10 times/sec
SLOW: 2 times/sec
<value> 1 to 255 in NR1
For example: WrtCmd (“APER MED, 55”)
Query syntax: APERture?
Return format:
FAST
MED , <NR1><NL^END>
SLOW