TH2829X Series Operation Manual Chapter 8 Performance and Test
Test frequency 100Hz 1kHz 10kHz 100kHz Test respectively
Level 1V
Range AUTO
Bias 0V
Speed Slow
Open and short correction should be made before testing. Connect standard AC
resistors: 10Ω, 100Ω, 1kΩ, 10kΩ, 100kΩ and change the frequency. The error
between reading and nominal value should be in the range ruled in this chapter.
8.6.10 Accuracy of DCR
Test condition
Function DCR
Test frequency -----
Level -----
Range AUTO
Bias -----
Speed Slow
Short correction should be made before testing. Connect standard DC resistors:
0.1Ω, 1Ω, 10Ω, 100Ω, 1kΩ, 10kΩ, 100kΩ. The error between reading and
nominal value should be in the range ruled in this chapter.