TH2829X Series Operation Manual Chapter 10 HANDLER
167
The FUNCtion:Source MONitor:VAC command is used to set the voltage monitor
ON or OFF. The FUNCTtion:Source MONitor:VAC? query returns the current
voltage monitor status.
ON (1)
FUNCtion:SMONitor:VAC OFF (0)
Where,
Character 1 (49) is equal to ON.
Character 0 (48) is equal to OFF.
For example: WrtCmd (FUNC:SMON:VAC ON) Set the voltage monitor as ON.
Query syntax: FUNCtion:SMONitor:VAC?
Return format: <NR1><NL^END>
The FUNCtion:SMONitor:IAC command is used to set the current monitor as ON
or OFF. The FUNCtion:SMONitor:IAC? query returns the current monitor status.
Command syntax:
ON (1)
FUNCtion:SMONitor:IAC OFF (0)
Where,
Character 1 (49) is equal to ON.
Character 0 (48) is equal to OFF.
For example: WrtCmd (FUNC:SMON:IAC ON) Set the current monitor as ON.
Query syntax: FUNCtion:SMONitior:IAC?
Return format: <NR1><NL^END>
The FUNCtion:DEV<n>:MODE command is used to set the deviation
measurement mode. The FUNCtion:DEV<n>:MODE? query returns the current
deviation measurement mode.
Command syntax:
ABSolute
FUNCtion:DEV<n>:MODE PERCent
OFF
Where,
ABSolute Absolute value deviation display
PERCent Percent deviation display
OFF Real value display
Where, <n> is
Character 1 (49) is equal to the nominal value of primary parameter.
Or character 2 (50) is equal to the nominal value of the secondary parameter.
For example: WrtCmd (FUNC:DEV1:MODE ABS)