TH2829X Series Operation Manual Chapter 10 HANDLER
181
Query syntax: CORRection:SPOT<n>:FREQuency?
Return format: <NR3><NL^END>
The CORRection:SPOT<n>:OPEN command is used to execute open correction
for specific frequency spots (frequency 1, frequency 2 frequency 201).
Command syntax: CORRection:SPOT<n>:OPEN
Where,
<n>:
Frequency spot 1~201
For example: WrtCmd (CORR:SPOT1:OPEN) Execute open correction for
frequency spot 1.
The CORRection:SPOT<n>:SHORt command is used to execute short correction
for specific frequency spots (frequency 1, frequency 2 frequency 201).
Command syntax: CORRection:SPOT<n>:SHORt
Where,
<n>:
Frequency spot 1~201
For example: WrtCmd (CORR:SPOT1:SHOR) Execute short correction for
frequency spot 1.
The CORRection:SPOT<n>:LOAD:STANdard command is used to set the
standard reference of specific frequency spots (frequency spot 1, frequency
spot2 frequency spot 201 which are taking load correction). The
CORRection:SPOT<n>:LOAD:STANdard query returns the current standard
reference of specific frequency spot.
Command syntax: CORRection:SPOT<n>:LOAD:STANdard <REF. A><REF. B>
Where,
<n>:
Frequency spot 1~201
<REF. A> can be NR1, NR2 or NR3 data format and taken as the standard
reference of the primary parameter.
<REF. B> can be NR1, NR2 or NR3 data format and taken as the standard
reference of the secondary parameter.
For example: WrtCmd (CORR:SPOT1:LOAD:STAN 100.7, 0.0002)
Query syntax: CORRection:SPOT<n>:LOAD:STANdard?
Return format: <NR3><NL^END>
The CORRection:USE:DATA? query returns the OPEN/SHORT/LOAD correction
measurement data of specific frequency spot 1, 2 201.