TH2829X Series Operation Manual Chapter 10 HANDLER
191
Command syntax: TRAN:Lk:LEVel<value>
Where,
<value> is NR1, NR2 or NR3 data format followed by mV or V.
Note: <value> ranges from 5mV to 2V. Beyond this range, error
information will be reported.
For example: WrtCmd (TRAN:Lk:LEVel 1V)
Query syntax: TRAN:Lk:LEVel?
Return format: <NR3><NL^END>
The TRAN:Lk:LIMit command is used to set the Lk nominal value, high and low
limits. The TRANS:Lk:LIMit? query returns the current Lk nominal value, high and
low limits.
Command syntax: TRAN:Lk:LIMit <value>, <low limit>, <high limit>
Where,
<value> is the Lk nominal value in NR1, NR2 or NR3 data format followed by H.
<low limit> is the Lk low limit in NR1, NR2 or NR3 data format.
<high limit> is the Lk high limit in NR1, NR2 or NR3 data format.
NOTE: The low limit should be smaller than the high limit, or error
information will be reported.
For example: WrtCmd (TRAN:Lk:LIMit 0.01H, -0.01H, 0.01)
Query syntax: TRAN:Lk:LIMit?
Return format: <NR3>,<NR3><NR3><NL^END>
The TRAN:Lx:STAT command is used to set the primary Lx ON or OFF. The
TRAN:Lx:STAT? query returns the current state of the primary Lx.
Command syntax:
ON
TRAN:Lx:STAT OFF
1
0
Where,
1 (decimal 49) is equal to ON.
0 (decimal 48) is equal to OFF.
For example: WrtCmd (TRAN:Lx:STAT:ON) Set all Lx parameters to be valid.
Query syntax: TRAN:Lx:STAT?
Return format: <NR1><NL^END>
The TRAN:Lx:FREQ command is used to set the Lx test frequency. The
TRAN:Lx;FREQ query returns the current Lx test frequency.
Command syntax: TRAN:Lx:FREQ <value>