EasyManua.ls Logo

Tonghui TH2829 Series - Page 208

Tonghui TH2829 Series
223 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
TH2829X Series Operation Manual Chapter 10 HANDLER
199
For example: WrtCmd (*ESR?)
The *STB? query returns contents of the standard service status byte register.
The execution of this command will not affect contents of the standard status
byte register.
Query syntax: *STB?
Return format: <value><NL^END>
Where,
<value> NR1 format: decimal expression for contents of the standard status
byte register.
Descriptions for each bit of the standard status byte register
For example: WrtCmd (*STB?)
The *OPC command equals to set the OPC bit of the standard event status
register when TH2829X finishes all parameter measurements. Ever since all
pending operations have been completed, this command will inform the
instrument to add a ASCII number 1 (decimal number: 49) into the output
buffer.
Command syntax: *OPC
For example: OUTPUT 717;“*OPC! Set the OPC bit of the instrument when
the last command is done.
Query syntax: *OPC?
Return format:<NL^END>
Where,
1 ASCII number 1decimal number: 49
For example: WrtCmd(*OPC?”)
Bit number
Description
7
6
5
4
3-0
Operation Status Register Summary Bit
RQS(Request Service) Bit
Standard Event Status Register Summary Bit
MAV(Message Available) Bit
Always 0(zero)