Program Mode 6-47
Fail End-of-Cycle Action
Choose from:
• Normal
• Auto Repeat
• Continuous Fill
• Continuous Fill Including Gross
Salvage End-of-Cycle Action
Choose from:
• Normal
• Auto Repeat
• Continuous Fill
The Continuous Fill Including Gross option found in the fail end-of-cycle actions
is only active after any step fails its limits, and used to continuously maintain air
pressure at the part as an aid in locating a leak. It is terminated by the Abort
signal.
If a user input is desired for an End-of-Cycle action, a Couple input can be used.
See "Coupling" on page 6-7