Quick Start 4-9
4.3.2 In Case of Difficulty
Many difficulties encountered when setting up a new tester are caused by simple
things. Address the items in the checklist below to ensure the tester is being operated
correctly.
Setup Checklist
• The key must be used to save changes made in Program mode. Be sure that it is
being used.
• Is the power supply connected to the back panel and is it receiving power (the
power supply’s green light is lit)?
• Is supply air connected to Qualitek mR?
• Is the supply air clean and dry? Air that contains moisture, oil, or hard
particulate will damage Qualitek mR.
• Is the supply air within the correct pressure range?
• Has the pressure regulator changed from when the test pressure was set? A
change will cause pressure errors.
• Has the security lock switch been set in the appropriate position for the desired
task (Run vs. Program mode)?
• Has the sensor calibration been changed?
• Are units set to the correct units of measure?
If the problem persists, call Uson for help.
Uson’s Customer Support Hotline is available at:
855-GET-USON (855-438-8766)
Changes made in Program Mode are NOT saved until the tester is returned to
Run Mode by using the KEY.