Reference 8-27
8.10.5 QmR Setup
The calculated offset and factors are now transferred to the QMR Calibration and
Temp Comp menus.
8.10.5.1 Comp Offset
If a calibration offset is required, select the Program Mode Test Setup Calibration
shown below. Set the Measure Setup sub tab Comp Method to Normal and enter
the Calibration Offset value into the Measure Setup sub tab Comp value
parameter.
8.10.5.2 Temp Comp Parameters.
Select the Calibration Temp Comp sub tab and set the TC Method to Enabled for
individual tested programs or the first program within a sequence as shown below.