<6. Parameter Lists>
174
IM 01E21A02-03EN
Relative
Index
Index Parameter Name Default Value
Write
Mode
Description
64 1064 FD_CHECK_PRI 0 AUTO
ShowstheFD_CHECK_ALMpriority
of the alarm. To activate an alarm
on communication, it is necessary
to specify the priority for alarm
activation in advance. To activate an
alarm, set 3 or more. By sending with
priority information being added in a
communication frame when the alarm is
activated, for example, the alarm is used
toapplyaltertoignorevaluesbelow
thespeciedpriorityonthehostside.
65 1065 FD_SIMULATE
0x00000000
0x00000000
1
AUTO
-
AUTO
Parameter to simulate an alarm.
66 1066
FD_
RECOMMEN_
ACT
0 -
Indicates countermeasure for the most
important alarm.
67 1067
FD_EXTENDED_
ACTIVE_1
0x00000000 -
Parameter to initiate an alarm. For
details about bit assignment, refer to FD
and bit items in the table in Subsection
5.6.2.
68 1068
FD_EXTENDED_
ACTIVE_2
0x00000000 -
69 1069
FD_EXTENDED_
ACTIVE_3
0x00000000 -
70 1070
FD_EXTENDED_
ACTIVE_4
0x00000000 -
71 1071
FD_EXTENDED_
ACTIVE_5
0x00000000 -
72 1072
FD_EXTENDED_
ACTIVE_6
0x00000000 -
73 1073
FD_EXTENDED_
ACTIVE_7
0x00000000 -
74 1074
FD_EXTENDED_
ACTIVE_8
0x00000000 -
75 1075
FD_EXTENDED_
MAP_1
0x000010E0 AUTO
Parameter set by the user as the mask
from FD_EXTENDED_ACTIVE_1 to
DEVICE CONDITION_ACTIVE_1.
76 1076
FD_EXTENDED_
MAP_2
0x7F047F0F AUTO
Parameter set by the user as the mask
from FD_EXTENDED_ACTIVE_2 to
DEVICE CONDITION_ACTIVE_2. For
details about bit assignment, refer to FD
and bit items in the table in Subsection
5.6.2.
77 1077
FD_EXTENDED_
MAP_3
0xE43D410B AUTO
Parameter set by the user as the mask
from FD_EXTENDED_ACTIVE_3 to
DEVICE CONDITION_ACTIVE_3. For
details about bit assignment, refer to FD
and bit items in the table in Subsection
5.6.2.
78 1078
FD_EXTENDED_
MAP_4
0xDF3F09FF AUTO
Parameter set by the user as the mask
from FD_EXTENDED_ACTIVE_4 to
DEVICE CONDITION_ACTIVE_4. For
details about bit assignment, refer to FD
and bit items in the table in Subsection
5.6.2.
79 1079
FD_EXTENDED_
MAP_5
0x00000400 AUTO
Parameter set by the user as the mask
from FD_EXTENDED_ACTIVE_5 to
DEVICE CONDITION_ACTIVE_5. For
details about bit assignment, refer to FD
and bit items in the table in Subsection
5.6.2.

Table of Contents

Other manuals for YOKOGAWA ADMAG TI Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA ADMAG TI Series and is the answer not in the manual?

YOKOGAWA ADMAG TI Series Specifications

General IconGeneral
BrandYOKOGAWA
ModelADMAG TI Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals