5-11
IM MW100-01E
Specification
5
File division: Files can be divided at arbitrary times using the event action
function.
Valid when the recording start action is Direct and recording stop
action is Fullstop or Rotate.
File message: The maximum number of characters that can be speci ed is 120.
The le message can be viewed when the le information is
displayed on the MW100 Viewer Software.
Write message: During execution of the recording action, a message that
corresponds to the recorded data can be included in the le.
No. of characters in message: Up to 15
Messages: 5
No. of write operations: Up to 10 per le
Operation upon failure recovery:
If a power failure occurs during recording, the data up to the
failure is restored upon recovery from the failure. The data after
recovery is recorded continuously to newly created les.
Operation when synchronizing time:
If time synchronization to SNTP is carried out during recording,
the time is saved to the recording log le.
Thinning Recording Function
Apart from the measured and computed values file, a set of thinned values from the
measured and computed values are saved to CF card.
Record start/stop: Executed simultaneously with the start/stop of the recording
of measured and computed values. No trigger functions are
available. User can select Thinning record or Do not record
Recording stop action: Select a record stop action of Single, Full stop, or Rotate.
See recording stop action in “Recording of Measured/Computed
Values”
Thinning time: Data saving is set for 1 per thinning time.
Set a thinning time of: 4, 5, 10, 20, or 30 sec; 1, 2, 3, 4, 5, 10, 20,
or 30 min.; or 1 hour. However, the thinning time cannot be set
shorter than the measurement interval.
Recording channels: Can be speci ed for each channel (settings for recording of
measured and computed values is set separately)
File name: Generated automatically in sequence using the date and serial
number (cannot be speci ed by the user).
Data length: Select 30 minutes, or 1, 2, 3, 4, 6, 8, or 12 hours, or 1, 2, 3, 5,
7, 10, 14, or 31 days. However, it cannot be set so that the le
size could exceed 10 Mbyte. Also, the data length cannot be set
shorter than the thinning time.
File size: The le size equation is the same as that for the computed value
les.
File division: Files can be divided at arbitrary times using the event action
function.
Valid when the recording stop action is Fullstop or Rotate.
File message: The maximum number of characters that can be speci ed is 120.
The le message can be viewed when the le information is
displayed on the MW100 Viewer Software.
Writing message: During execution of the recording action, a message that
corresponds to the recorded data can be included in the le. Five
messages of up to 15 characters each are available for including
in a single le, up to 10 messages per le.
5.2 Main Module Specifications

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA MW100 and is the answer not in the manual?

YOKOGAWA MW100 Specifications

General IconGeneral
BrandYOKOGAWA
ModelMW100
CategoryData Loggers
LanguageEnglish

Related product manuals