Error Messages
MT8852B Operation Manual PN: 13000-00205 Rev. N D-3
EUT test mode rejected On completion of the ACL connection, the MT8852B sends
an activate test mode command to the EUT. The EUT must
have been preconfigured to accept this command. If the EUT
has not been initialized to go into test mode, this error
message is shown. Only Null packet mode testing can be
performed in this condition.
Error decoding the packet type Decoded packet type does not match the correct type.
Error finding packet start There was no packet burst found.
Error locating EDR sync word The EDR synchronization word could not be located. The
location of the EDR synchronization word is incorrect or the
EDR synchronization word is incorrect/corrupted.
Error locating GFSK sync word The GFSK synchronization word could not be located. A
common cause is the access code is set incorrectly or is
corrupted, or the location of the GFSK synchronization word
is incorrect.
Error!< Already in RUN The RUN command has been sent to the MT8852B but the
MT8852B is already RUNning a test/script.
Error!< BLE Addr Srce Invalid The EUT address source is set to Inquiry or Manual and BLE
test cases are also selected. For BLE test cases, the valid
EUT address sources are RS232, USB or BLE2WIRE.
Error!< BT Addr Srce Invalid The EUT address source is set to BLE2WIRE and BR/EDR
test cases are also selected. BLE2WIRE is only valid for BLE
test cases.
Error!< EUT in Test Mode The MT8852B has sent the Enable Test mode command to
the EUT but the EUT is already in test mode. The enable test
mode command needs to be sent only once to a EUT
Error!< External HCI The MT8852B has been configured to be controlled by an
external device through its rear panel RS232 HCI interface.
To return to normal operating mode, press
CONFIG\MT8852B\System interfaces – and set “RS232
mode” to EXT COMMS.
Error!< GPIB test config When an instrument is in single test mode and configuration
change or data read is attempted for another test, this
message will appear.
Error!< In chan display mode The MT8852B is in channel display mode but a frequency
command was sent to the MT8852B. When in channel
display mode, only channel commands are valid.
Error!< Invalid BLE Frequency An invalid BLE frequency was sent during the configuration of
a BLE test case. BLE frequencies are spaced 2MHz apart.
Error!< Num packets not even For the BLE PER test case, an odd number of packets have
been sent to the MT8852B. For this test case, the number of
packets must be an even number.