Error Messages
D-4 PN: 13000-00205 Rev. N MT8852B Operation Manual
Error!< Out of range value A test case configuration parameter was sent to the
MT8852B that is outside of the valid parameter range. Check
the parameters being sent to the MT8852B. Refer to the
measurement setup pages of the manual for valid ranges.
Error!< Password required The script that is being edited is locked by a password. Enter
the password to edit the script.
Error!< Slave mode The MT8852B is in 'Make me a EUT mode' and an
incompatible request has been sent to the MT8852B. Preset
or Restart the MT8852B to reset the state of the MT8852B.
Error!< in calibration mode This message is output when trying to perform a test run
whilst the instrument is in calibration mode.
Error!< locked script It is not possible to edit a script that has been locked by
another user. To lock and unlock scripts, press Scpt/Test,
move the pointer to the script number and press the List soft
key. The status of all scripts is then shown and can be
changed using the soft keys.
Error!< non-editable script Scripts 1 and 2 are factory set to represent common test
requirements and cannot be altered by the user. To define
your own script settings use scripts 3 to 10.
External HCI interface The instrument’s Bluetooth core is now driven only by the
RS232 connection on the MT8852B back panel using
standard HCI commands.
GFSK sync word exceeds limits The GFSK synchronization word has been detected outside
of its expected location.
Hardware failure The MT8852B has reported a Hardware error. Restart the
MT8852B.
Host timeout The EUT has not responded to a connection request before
the accept connection timer has expired.
IQ Modem DAC Saturation The MT8852B IQ modem has reported an error. Check the
EUT transmit power level and re-RUN the test.
Incorrect packet length The packet received from the EUT is not the full length of a
standard DH1, DH3 or DH5 packet. It is therefore not suitable
for measurements.
Internal sync error, ET busy The MT8852B has reported a synchronization error. Restart
the MT8852B.
Invalid Air code When attempting to establish a SCO connection with the
EUT, the MT8852B air code must correspond to the air code
used by the EUT. Press Config \ MT8852B \ Audio and set
the air code format to match that of the EUT, before
attempting a SCO connection.
Invalid LMP parameters The EUT rejected the LMP due to the parameters not being
compatible with its present state.
Table D-1. Error Messages
Message Meaning