Running a Script in Standard Mode Basic Rate Tests
MT8852B Operation Manual PN: 13000-00205 Rev. N 7-17
FER due to length: Displays the number of packets returned with incorrect
payload lengths. This will also result in a CRC error. These
packets are used in BER calculations.
FER due to lost packets:A lost packet is counted by the MT8852B under the following
conditions:
Not recognizing the sync word header as being correct.
The packet failing the Header Error Correction (HEC)
No packet received during the expected time slot.
NACK packet is received.
These packets are not used in BER calculations.
5. Press the [Extended] soft key to display more detailed test results. An example of the
first extended page for the single sensitivity test is shown below.
6. Page through the extended screens with the soft key in the normal manner, or press the
[Summary] soft key to revert to the initial summary page.
Figure 7-23. Single Sensitivity Extended Results