288 | Appendix A: Error Codes
Code Message What to do
1301 Syringe module not connected Make sure the Syringe module is correctly
connected using the new BioTek-provided
serial cable.
1302 Syringe initialization error
1303 Syringe sensor not cleared
error
1304 Invalid syringe dispense volume See error 1308 below.
1305 Invalid syringe operation Contact BioTek TAC.
1306 FMEA error on syringe A Contact BioTek TAC.
1307 FMEA error on syringe B Contact BioTek TAC.
1308 Invalid Syringe pre-dispense
volume
Protocol may have been written for a
different type of dispense manifold. Make
sure the Instrument Settings represent the
installed hardware. Modify the protocol to
match.
1309 Invalid Syringe prime volume
1310 Invalid Syringe manifold
1350 Peri-pump (PP) invalid dispense
volume
Contact BioTek TAC.
1351 PP invalid cassette type Change the cassette type to match the
protocol requirement and rerun the protocol.
1352 PP invalid pre-dispense volume Contact BioTek TAC.
1353 Multiple required cassettes Make sure every Peri-pump step in the
protocol calls for the same cassette type. A
conflict was found.
1354 PP pump cover (safety door)
open
Close the pump cover door and rerun the
protocol.
1355 PP not installed If you are trying to run a Secondary Peri-
pump, check the cabling in the back of the
instrument. Otherwise Contact BioTek TAC.
1356 PP Invalid dispense position Contact BioTek TAC.
1357 PP two dispensers are expected Contact BioTek TAC.
1358 PP Half UL not supported 0.5 µL Peri-pump dispensing is not
supported on this instrument.